![FujiFilm PB2020-M Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/fujifilm/pb2020-m/pb2020-m_operation-manual_2343363046.webp)
(6) Using a syringe, flush 60 mL of disinfectant
solution onto each gap and recess of the probe
connector to remove air bubbles completely.
(7) Keep the endoscopic ultrasonic probe immersed
in the disinfectant solution at the temperature,
concentration and time recommended by the
disinfectant solution's manufacturer. Cover the
basin with a lid during disinfection.
(8) Remove the endoscopic ultrasonic probe from
disinfectant solution.
(9) Wipe any disinfectant solution remaining on the
external surface of the endoscopic ultrasonic
probe with sterile gauze.
202B1263042
46
Chapter 6 Chemical Disinfection
PB2020-M_E2-10_202B1263042.indb 46
2016/03/16 16:43:39