4.2
Withdrawal of Endoscopic Ultrasonic Probe
WARNING
Do not withdraw this product sharply. In so doing, body fluids adhered to
this product may be splattered, posing an infection risk.
CAUTION
When withdrawing this product, be sure to press the [FREEZE] key on
the ultrasonic processor to enable freeze mode. Withdrawing this product
while it is in operation may cause damage to it.
When this product is used with an endoscope with a forceps stand
mechanism, make sure that the forceps stand is in neutral position before
withdrawing this product. Withdrawing this product with the forceps stand
raised may cause damage to it.
If it is difficult to remove this product from an endoscope, slowly withdraw
it while monitoring an endoscopic image in order not to damage body
cavities.
(1) Unlock the bending section by turning the
up-down and left-right angulation locks.
(2) Operate the up/down and left/right angulation
knobs to straighten the bending section to its
neutral position.
(3) Press the [FREEZE] key on the keyboard of the
ultrasonic processor to enable freeze mode.
[Note]
Make sure that the transducer of the ultrasound
head stops rotating.
202B1263042
33
Chapter 4 Method of Use