7.2
Ethylene Oxide Gas Sterilization
WARNING
When performing ethylene oxide gas sterilization of this product,
ensure that the surfaces are dry before attempting ethylene oxide gas
sterilization. If any moisture remains in this product, sterilization of undried
parts will be incomplete and it could be a source of infection.
When performing ethylene oxide gas sterilization of this product, detach
the connector cap from the probe connector. If sterilization is performed
while the connector cap is placed on the probe connector, sterilization will
be incomplete and may pose an infection risk.
This product must be properly aerated after ethylene oxide gas
sterilization to remove toxic components. Gas remaining in this product
after ethylene oxide gas sterilization may harm human body.
CAUTION
In order to successfully sterilize this product by ethylene oxide gas
sterilization, use the parameters described in “Table 7.2 Conditions of
20% ethylene oxide gas sterilization (chamber type).” Exceeding the
recommended conditions may cause equipment damage.
Table 7.2 Conditions of 20% ethylene oxide gas sterilization (chamber type)
Ethylene oxide gas sterilization
Aeration
Temperature Humidity
Pressure
Time
Temperature
Time
55ºC
Normal
humidity
100 kPa or more
(relative pressure)
4 hours
55ºC
12 hours
202B1263042
50
Chapter 7 Sterilization