4-24
4 Applicable usage methods
897N201058J FDC NX700 Instruction Manual 2022.04
4. Load the sample.
(1) Place the specified amount of whole blood
sample in a blood collecting tube.
NOTE
The PF has about 3 mL of dead vol-
ume (that will be absorbed into the
filter). If the sample volume is low,
the filtered sample volume will be
too low for measurement. Further-
more, sample that was aspirated
into the PF filter cannot be reused.
NOTE
If the sample volume is too high,
the sample may overflow from the
blood collecting tube when the PF is
loaded. There must be empty space
in the tube. (Refer to Section 4.4.2.)
(2) Mix the whole blood sample in the blood
collecting tube by turning it over five or six
times.
NOTE
Make sure that bubbles do not form.
(3) Remove the cap from the blood collecting
tube, and then put it on the PF sample rack.
(4) Insert a PF into the blood collecting tube.
Make sure the PF does not rise from the
sample rack.
START
STOP
0
3
A
B
D
C
E
MYID001
..0160503010
..0160503009
5. Tap a sample rack display. In the Sample
information registration screen, register
the measurement information, and then
tap the [OK] button.
IMPORTANT:
Tests using a PF measure filtered plasma, so al-
ways configure the sample type to [Plasma/Serum].
4
Содержание FUJI DRI-CHEM NX700
Страница 2: ......
Страница 22: ......
Страница 72: ......
Страница 114: ......
Страница 126: ......
Страница 166: ......
Страница 173: ...8 Troubleshooting 8 3 897N201058J FDC NX700 Instruction Manual 2022 04 8 ...
Страница 216: ......
Страница 258: ......
Страница 262: ...11 USM User Safety Mode 11 4 897N201058J FDC NX700 Instruction Manual 2022 04 ADM All functions are operable 11 ...
Страница 272: ......
Страница 274: ......
Страница 275: ......