6-33
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
If the check results are [NG], tap the [Cleaning]
button, clean the PF unit, and then run the leak
test again.
If this does not resolve the problem, tap the
[Procedure] button, and then check the attach-
ment or replace the PF pump tube to using the
procedure on the following page.
NOTE
If the [Result details] button is tapped,
the Advanced leak check results display
screen appears.
NOTE
The check results can be printed by tap-
ping the
button.
7. Once the check ends, remove the PF
sample rack and PF leak check tool.