8 Troubleshooting
8-17
897N201058J FDC NX700 Instruction Manual 2022.04
(8) During a diluent / reference fluid error
Error
code
Error message
Error condition
Troubleshooting
E0121
Aspiration clogging
ERR (diluent)
Clogging was detected
during sampling.
Check the diluent.
Clogging was detected
during diluent aspira-
tion.
Check whether the diluent / reference
fluid are loaded in a proper blood col
-
lecting tube or tube. If there was a load-
ing problem, replace the diluent and
reference fluid, and then rerun from the
measurement request.
NOTE
If the error occurs again, there
might be a piping system prob-
lem. Please contact your dealer
for assistance.
E0122
Spotting clogging ERR
(diluent)
Clogging was detected
during sampling.
Check the diluent.
Clogging was detected
during diluent applica-
tion.
E0127
Aspiration clogging
ERR (diluent)
Clogging was detected
during sampling.
Check the sample and
diluent.
Clogging was detected
during reference fluid
aspiration.
E0128
Spotting clogging ERR
(diluent)
Clogging was detected
during sampling.
Check the sample and
diluent.
Clogging was detected
during reference fluid
application.
8
Содержание FUJI DRI-CHEM NX700
Страница 2: ......
Страница 22: ......
Страница 72: ......
Страница 114: ......
Страница 126: ......
Страница 166: ......
Страница 173: ...8 Troubleshooting 8 3 897N201058J FDC NX700 Instruction Manual 2022 04 8 ...
Страница 216: ......
Страница 258: ......
Страница 262: ...11 USM User Safety Mode 11 4 897N201058J FDC NX700 Instruction Manual 2022 04 ADM All functions are operable 11 ...
Страница 272: ......
Страница 274: ......
Страница 275: ......