8 Troubleshooting
8-42
897N201058J FDC NX700 Instruction Manual 2022.04
(5) Standard curve configuration error
Error
code
Error message
Error condition
Troubleshooting
E4202
Calibration curve set-
ting ERR (span error)
Misplacement of cali-
brator.
Check the level and
expiration date of the
calibrator and perform
the calibration from the
beginning.
The wrong calibrator
may have been used.
Check the expiration date and the set-
ting position of the calibrator, and then
calibrate again from the beginning.
E4203
Calibration curve set-
ting ERR (range error)
Misplacement of cali-
brator.
Check the level and
expiration date of the
calibrator and perform
the calibration from the
beginning.
E4205
Calibration calcula-
tion ERR
Failed with error in
calibration operation.
Do the calibration
again.
Calibration calculation
problem
Calibrate again from the beginning.
(6) Measurement errors during calibration
Error
code
Error message
Error condition
Troubleshooting
E4204
Calibration measure-
ment ERR
Error in measured
result.
After coping with the
printed error, perform
the calibration from the
beginning.
Error marks such as
“ * ”, “ ? ”, and “$” are
printed on the meas-
urement results.
Troubleshoot the printed errors and
then calibrate again from the beginning.
8
Содержание FUJI DRI-CHEM NX700
Страница 2: ......
Страница 22: ......
Страница 72: ......
Страница 114: ......
Страница 126: ......
Страница 166: ......
Страница 173: ...8 Troubleshooting 8 3 897N201058J FDC NX700 Instruction Manual 2022 04 8 ...
Страница 216: ......
Страница 258: ......
Страница 262: ...11 USM User Safety Mode 11 4 897N201058J FDC NX700 Instruction Manual 2022 04 ADM All functions are operable 11 ...
Страница 272: ......
Страница 274: ......
Страница 275: ......