8 Troubleshooting
8-18
897N201058J FDC NX700 Instruction Manual 2022.04
(9) During a calibrator volume error
Error
code
Error message
Error condition
Troubleshooting
W0061
Calibrator shortage
Calibrator is insuffi
-
cient.
Check the calibrator
volume and perform
calibration or verify
calibration again.
There is not enough
calibrator.
Check the calibrator liquid volume, and
then measure again.
W0145
Insufficient calibrator
Calibrator is insuffi
-
cient.
Check the calibrator
volume and perform
calibration or verify
calibration again.
W0146
Surface detection ERR
(CP)
Calibrator volume is
too much.
Check the calibrator
volume and perform
calibration or verify
calibration again.
There is too much cali-
brator.
Make sure that the cap of the calibrator
has been removed and that there is no
liquid film in the opening section, and
then rerun the measurements.