6-19
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
2
3
1
Connector
Thumbscrew A
Thumbscrew B
6. Insert the connector of the light source
lamp.
7. Insert the lamp unit into the analyzer
with thumbscrew B facing up. Tighten
thumbscrew A securely.
IMPORTANT:
Tighten the thumbscrew securely. Otherwise, the
measurement results may be negatively affected.
8. Attach the front bottom cover, load the
disposal box, and then switch on the
power of the FDC NX700.
IMPORTANT:
Be sure to load the disposal box. If the analyzer
is used without the box, the measurement results
may be negatively affected.
9. After starting the analyzer, go to [Lamp
management] in Normal mode, and then
reset [Lamp ON time].