6-10
6 Maintenance
897N201058J FDC NX700 Instruction Manual 2022.04
Transfer bar
7. Use gauze or a cotton swab that has
been soaked in lukewarm or cold water
to wipe the upper surface of the transfer
bar visible in the sample application unit.
8. Next, clean the incubator, sample appli-
cation unit, and ISE test unit.
NOTE
After cleaning, attach the removed slide
transfer weight and sample application
unit cover at their original positions, and
then tap the button to return to the
previous screen.
CAUTION:
Keep your fingers clear as the sample disk moves.
9. Tap the
button to switch off the
power.
CAUTION:
Keep your fingers clear as the sample disk moves.
6