6.9.3 Replacing the PF Suction Pad
Replace the PF Suction Pad once a year.
WARNING:
Process used PF Suction Pad as infectious waste in compliance with any applicable regulations in
your country.
Tips
1. With [Warming up], [Lamp warming up]
or [Ready to test] displayed on the LCD,
make sure that the analyzer is not apply-
ing samples.
2. Tap the tip display on the LCD.
3. The Tip loading screen appears.
The sample disk turns, and the consumables
are moved to the front.
CAUTION:
Do not touch the sample disk until it stops.