202B1262575D
Problem
Cause
Remedy
Distorted image
1) The imaging section is damaged.
2) The ultrasonic endoscope is not connected
correctly to the processor.
3) The video signal cable has shorted out or
broken.
1) Reset
[Note]
the processor and the light
source. If an appropriate image does not
appear even after resetting, turn them off,
straighten the bending section to unlock,
release the angulation knobs, and then
withdraw the ultrasonic endoscope slowly
from the patient.
2) Connect the ultrasonic endoscope properly
to the processor.
3) Reset
[Note]
the processor and the light
source. If an appropriate image does
not appear even after resetting and it is
impossible to continue the procedure, turn
them off, straighten the bending section to
unlock, release the angulation knobs, and
then withdraw the ultrasonic endoscope
slowly from the patient.
No air and/or water
delivery
1) Pump is switched off.
2) Abnormality in the air/water button.
3) The air/water button is not pressed firmly
while supplying sterile water.
4) Water tank cap is loose.
5) Water tank is filled with too much sterile
water.
6) Water tank is empty.
7) Water tank is not connected.
8) Clogged nozzle or channel.
1) Switch on the pump.
2) Replace with a new air/water button.
3) Push in the air/water button firmly.
4) Close the cap firmly.
5) Reduce the water level in the water tank to
about 80% of its capacity.
6) Fill the water tank with sterile water.
7) Connect the water tank.
8) Clean clogged channel per provided
instructions.
Low air/water supply
amount
1) Foreign matters have adhered to the air/
water channels.
2) The air/water channels are damaged.
1) After pulling out the ultrasonic endoscope
from a patient, clean the air/water channels
according to the procedures specified
in the Reprocessing Manual “Chapter 3
Pre-cleaning” and “Chapter 4 Manual
Cleaning”.
If air/water supply amount is still low,
replace with a spare ultrasonic endoscope.
2) Contact your local FUJIFILM dealer.
[Note] To reset the processor and the light source, turn them off, and wait for at least 5 seconds. Turn on the processor and
the light source again, and then light the lamp by pressing the Lamp button.
[Note] To reset the ultrasonic processor, turn it off, and wait for at least 5 seconds. Turn on the ultrasonic processor again.
Chapter 5 Troubleshooting
95
EG580UT_E2-50_202B1262575D.indb 95
2016/07/05 9:28:56