202B1262575D
(6) Insert the puncture needle into the ultrasonic endoscope.
(7) Stop inserting the puncture needle when its tip appears in
the endoscopic image.
(8) Bring the puncture needle close to the target area slowly.
[Note]
The puncture needle may remain bent if the ultrasonic
endoscope is angulated with the puncture needle inserted.
(9) Perform puncture.
[Note]
The puncture guiding line is only meant as a guide. Check
the actual part to be punctured on the ultrasonic image.
[Note]
The puncture needle may remain bent if puncture is
performed while the bending section is bent.
[Note]
The puncture needle may remain bent if puncture is
performed while the forceps stand is positioned upright.
Chapter 4 Method of Use
86