Main Specification
<Classification of Medical Electrical Equipment>
1. Type of protection against electric shock : Class I equipment (Power supply: Protective
earth plug)
2. Degree of protection against electric shock : Type BF applied part
3. Degree of explosion protection
: Use is prohibited in an oxygen-rich
environment or in a flammable gas
atmosphere.
4. Degree of waterproof
: IEC 60529 IPX7 (with US waterproof cap
attached)
[Note] Use in combination with the VP-4450HD processor, XL-4450 light source and
SU-1 ultrasonic processor.
<Applied Part>
Insertion portion
<Data about Main Unit>
Endoscopic functions
Model
EG-580UT
Viewing direction
40°
Observation range
3 to 100 mm
Field of view
140°
Distal end diameter
13.9 mm
Maximum diameter of insertion portion
15.0 mm
Insertion tube diameter
12.4 mm
Flex
[Note 1]
10.5 N
Bending capability
UP/DOWN
150°/150°
LEFT/RIGHT
120°/120°
Minimum diameter of instrument channel
3.8 mm
[Note 2]
Working length
[Note 3] [Note 4]
1250 mm
Overall length
1550 mm
Insertion route
Peroral
[Note 1] Shows reaction force for the area of 200 mm from the distal end.
[Note 2] It does not mean that compatibility of machines selected based on only this
instrument channel diameter is guaranteed.
[Note 3] Use an endotherapy device with a working length of 1800 mm or longer.
[Note 4] As a puncture needle, use an aspiration biopsy needle for ultrasonic endoscopes with
a working length of 1400 mm and for an instrument channel diameter of 3.8 mm.