202B1262575D
Electromagnetic immunity compliance information and guidance
Immunity test
EN 60601-1-2
Test level
Compliance
level
Guidance
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
3V[V
1
]
3V/m[E
1
]
Portable and mobile RF communications equipment
should be used no closer to any part of this product,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of transmitter.
Recommended separation distance
d=
√
P
V
1
3.5
d=
√
P
E
1
3.5
80 to 800MHz
d=
√
P
E
1
7
800MHz to 2.5GHz
Where “P” is the maximum output power rating of
transmitter in watts (W) according to the transmitter
manufacturer and “d” is the recommended separation
distance in meters (m).
This product complies with the requirements of
EN 60601-1-2: 2007.
However electromagnetic interference may occur on this
product under electromagnetic environment that exceeds
its noise level.
Electromagnetic interference may occur in the vicinity of
equipment marked with the following symbol.
Electromagnetic immunity compliance information and guidance
The customer or the user of this product can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitter) and this product as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter P (W)
Separation distance related to frequency of transmitter (m)
150kHz to 80MHz
d=1.2
√
P
80 to 800MHz
d=1.2
√
P
800MHz to 2.5GHz
d=2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
107
Main Specification
EG580UT_E2-50_202B1262575D.indb 107
2016/07/05 9:28:57