Chapter 3 Preparation and Inspection Before Use
CAUTION
Do not use the ultrasonic endoscope if the inspection result shows
any abnormality. Doing so can negatively affect the functionality of the
instrument or increase risks to patient safety.
3.1 Preparing Forceps Valve
WARNING
Ensure that the forceps valve is properly attached to the instrument
channel inlet. If this product is used without the forceps valve attached,
body fluid may flow back and it could be a source of infection.
Attach a properly disinfected (or sterilized) forceps valve. If the forceps
valve is not cleaned or disinfected (or sterilized) properly, it can create
a risk of infection to patients and/or end-users.
Be sure to inspect the forceps valve before attachment. If any
abnormality is found during inspection, do not use the product. It could
be a source of infection.
3.1.1 Cleaning and Disinfecting (or Sterilizing) the Forceps Valve
The forceps valve is provided without sterilization and must be cleaned and disinfected (or
sterilized) prior to use following the instructions of the Reprocessing Manual.
The forceps valve is a single-use product. Do not reuse it to avoid infection.
[Note] For information on how to clean and disinfect (or sterilize) the forceps valve, refer
to the Reprocessing Manual.
Reprocessing Manual “7.1 Cleaning and Disinfection (or Sterilization) of Forceps
Valve”