56
INZ-TN2ZRE-E
<How to set/change the range>
The measuring range can be arbitrarily selected in
the minimum and the maximum range specified at
the time of purchase. The range to be used can be
selected 1 or 2.
(1) Move the cursor to the item to be set by press-
ing the or the key, and then press the
ENT
key.
ENT
(2) Move the cursor to the Ch (component) whose
setting is to be changed by pressing the or
the key, and then press the
ENT
key.
(3) Move the cursor to the item whose setting is to
be changed by pressing the or the key,
and then press the
ENT
key.
(4) Press the or the key to change the
value. Press the key to select the next digit.
The unit cannot be changed.
In a state where the decimal point is highlight-
ed, press the or the key, and the decimal
point position can be changed.
(5) When necessary change is made, press the
ENT
key.
The value for range 1 and range 2 must fall within
the range from the MIN and the MAX range
(including the MIN and the MAX range), and
at the same time range 1 must be smaller than
range 2.
The number of ranges is 1 or 2.
Settable range
ENT
ENT