49
INZ-TN2ZRE-E
ENT
(5)
The value is highlighted, indicating that
the value can be changed. Change the
value by pressing the or the key,
and then move the cursor to the right digit
by pressing the key.
(6)
After the value is changed, press the
ENT
key.
ENT
End of Hold setting
Parameter setting screen
ESC
The setting is expressed in % against the range
for both ranges.
When 0 to 1000 ppm is selected as the range,
for example, if 10% FS is selected as hold
setting, the output equivalent to 100 ppm is
output and held irrespective of the measure-
ment value at that time.
Meaning of setting
• Instantaneous value display of the measurement cannot be held. (Output only can be held.)
• If set value is selected for hold, instantaneous O
2
correction value is calculated and held
based on the set value.
• Range identification contact output cannot be switched even if the range is switched during
the hold.
Description of setting
(7)
Press the key to return to the param-
eter setting screen.
ESC
Response time
The response time of the electrical system can
be changed.
Setting is available by components.
Note) It does not provide exact seconds for
the setting time, but it gives a guide
of the setting time.
The setting value can be modified as
requested by the customer.