6
INZ-TN2ZRE-E
3.2 installation
3.2.1 installation of analyzer main frame
Installation methods for the analyzer main unit are divided into 2 types;
Note) • The analyzer weight should be supported at the bottom of the casing.
• The analyzer should be installed in a place where ambient temperature is within -5 to 45°C (max.
40°C when two optical units are used, and the power supply is more than 200V AC), and tem-
perature fluctuation during using is minimum.
• Where vibration is unavoidable, protect the analyzer from vibrating.
For example, protection rubber is installed between the analyzer and support, or the analyzer’s
front panel and mounting panel.
“A” : 57.2 (EIA) or 50 (JIS)
“A” : 57.2 (EIA) or 50 (JIS)
External dimensions
Mounting dimensions
Mounting method
(Unit : mm)
Type
19-inch rack
mounting
Panel cutout dimensions
Panel mounting
ESC
SPAN
MODE
ENTER
ZERO
GAS
ZRE
ANALYZER
ESC
SPAN
MODE
ENTER
ZERO
GAS
ZRE
ANALYZER
436
+2
-0
126
+2 -0
Mounting bracket
Support
Support
450 or more
465
A
M6
435
483
443
132.5
132.5
A