If calibration is cancelled after the calibration gas is supplied regardless of during manual/auto
calibration, the holding extending time will be performed.
6
You can select the value for hold from the value immediately before entering output hold, “cur-
rent,” and arbitrary value, “setting.”
Follow the procedures shown below to make the setting.
(1)
In the “Parameter setting” screen that ap-
pears, select “Output Hold”.
“ON” or “OFF” is highlighted by press-
ing the
ENT
key. Press the or the
key to select ON/OFF. Press the
ENT
key
to return to (1).
(2)
Where ON/OFF is highlighted, press the
key. “Current” or “Setting” is high-
lighted. Select “Current” or “Setting” by
pressing the or the key.
(3)
Press the
ENT
key while “Current” is se-
lected to return to (1). Press the
ENT
key
while “Setting” is selected to go to the
parameter hold screen.
“Current”: Holds the value immediately
before the hold.
“Setting”: Holds the value arbitrarily set.
ENT
(4)
On the parameter hold screen that ap-
pears, move the cursor next to the Ch
(component) you want to make the set-
ting by pressing the or the key,
and then press the
ENT
key.
ENT