INZ-TN2ZRE-E
• Output adjustment screen
Analog output adjustment screen.
Connect the digital multi meter to the output
terminal corresponding to the number of OUT
to be adjusted, and adjust the value so that 4mA
or 0V is output at zero and 20mA or 1V is
output at span.
Description of output adjustment screen
Move the cursor using the , , or
the key to the output (OUT No. and
zero/span) to be adjusted, and then press
the
ENT
key.
The selected value is highlighted. Adjust
the value, while watching the output, by
pressing the or the key. Press the
key to select the next digit.
On completion of the adjustment, press
the
ENT
key.
ENT
Be sure to perform zero / span calibration when the range setting is changed.
Otherwise, the measurement value may not be output properly.
Caution