6.4 setting of auto calibration
6.4.1 Auto calibration
Auto calibration is automatically carried out at the time when zero span calibration are set. Before
changing the setting of auto calibration, set the ON/OFF to OFF.
(1) Enter the “Setting of Auto Calibration”
screen from the user mode, and the
display shown at right appears. Operate
the or the key until the cursor is
aligned with a desired item and press the
ENT
key.
(2) In the “Setting of Auto Calibration”
screen that appears, perform the value
entry or the setting. For the value entry
or setting change, use the or the
key, and the key to move the cursor to
the right.
After setting, press the
ENT
key, and auto
calibration is carried out by the entered
setting value.
Description of setting items
• Start Time : Setting at the first calibration
(day of the week, hour, minute)
• Cycle
: A period between the start time of one
calibration and another
(unit : hour/day)
• Flow Time : The time required for replacement by
calibration gas
Time required for replacement of sample
gas after the calibration is completed
(Set by calibration gas. See the next
page.)
• ON/OFF
: ON/OFF of auto calibration
( )
ENT
ENT
To close the "Setting of Auto calibration" or cancel
this mode midway, press the
ESC
key.
A previous screen will return.
To close "Setting of Auto calibration"
ESC
End of Auto Calibration setting
Press the or
the key, and
date and time are
displayed
alternately.