![Fuji Electric ZKJ-2 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/fuji-electric/zkj-2/zkj-2_instruction-manual_2341904008.webp)
1 - 1
INZ-TN2ZKJ-E
1. OVERVIEW
This infrared gas analyzer (type: ZKJ) measures the concentration of NO, SO
2
, CO
2
, CO and CH
4
con-
tained in sampling gas on the principle that different atomic molecules have an absorption spectrum in the
wave band of infrared rays, and the intensity of absorption is determined by the Lambert-Beer law.
Since this instrument incorporates a compact paramagnetic O
2
sensor, it allows measuring up to 5 compo-
nents simultaneously by using the built-in O
2
sensor (up to 4 components if O
2
sensor is excluded).
Furthermore, use of a microprocessor or large sized liquid crystal display realizes improvement of oper-
ability, accuracy and multi-functions.
This instrument is optimum for measuring combustible gas exhausted from boilers or incinerators, and it
is effective for steel gas analysis (blast furnace, steel converter, thermal treatment furnace, sintering
(Pellet equipment), coke furnace), storage and maturity of vegetable and fruit, biochemistry (microbe),
[fermentation], air pollution [incinerator, exhaust gas desulfurization, denitration], automotive emission
(excluding tester), protection against disasters [detection of explosive gas and toxic gas, combustion gas
analysis of new building material], growth of plants, chemical analysis [petroleum refinery plant, petro-
leum chemistry plant, gas generation plant], environment [landing concentration, tunnel concentration,
parking lot, building management] and various physical and chemical experiments.