6 - 1
INZ-TN2ZKJ-E
6. SETTING AND CALIBRATION
6.1 Changeover of range
This mode is used to select the ranges of mea-
sured components.
q
During measurement, press the
MODE
key
to display the User mode.
w
Point the cursor to “Changeover of
Range”. Press the
ENT
key.
e
The “Channel Selection” screen appears.
Press the
or
key until the “ ”
cursor moved selects a desired CH
(component).
r
After selection, press the
ENT
key.
Note) The range of O
2
correction instanta-
neous values and O
2
correction
average values is automatically
switched by changing the range of
instantaneous value of each CH
(component).
t
In the Range Setting screen that appears,
move the cursor by pressing the
or
key to select the range. (The range
with a mark of is currently selected.)
y
After selection, press the
ENT
key.
u
Measurement is conducted within the
selected range. The range identification
contact is closed with the low range
(Range 1), and open with the high range
(Range 2).
Note) If the Remote Range is set to ON,
the changeover of range cannot be
performed on the screen.
MODE
(
)
ENT
(
)
ENT
(
)
ENT
To close Changeover of Range, or cancel
the command midway, press the ESC key.
A previous screen will return.
To close “Changeover of Range”
ESC
End of Range Selection