6 - 3
INZ-TN2ZKJ-E
t
In the “Calibration Concentration Selection”
screen that appears, select any concentration
item you want to set by pressing the
,
key.
y
In the “Calibration Concentration Value
Setting” screen that appears, enter calibration
gas concentration values (zero and span).
For value entry, press the
or
key, and
a 1-digit value increases or decreases. By
pressing the
, the digit moves.
After setting, save the entry by pressing the
ENT
key. The saved value becomes valid
from the next calibration process.
Note) Enter the set values corresponding to
each range. When the O
2
measurement
uses atmospheric for the zero gas, set
the concentration value to 20.60. When
the cylinder air is used, set to the con-
centration value as indicated on the
cylinder.
(
)
ENT
ENT
End of Calibration
Concentration Setting
To close the calibration concentration value
setting process or cancel this mode midway,
press the ESC key.
A previous screen will return.
To close the setting
ESC
NOx, SO
2
, CO
2,
CO, CH
4
, external O
2
measurement and buit-in paramagnetic O
2
sensor
Span gas: 1 to 100% of full scale
(Full scale (FS) is the same as each range
value.)
External Zirconia O
2
measurement
The setting cannot be performed beyond the range.
Zero gas: 5 to 25 vol%
Span gas: 0.01 to 5 vol%
Setting range of values
Cursor for setting value