6 - 14
INZ-TN2ZKJ-E
“Auto calibration status” and “maintenance status” contact outputs are closed during auto calibration, and
are open in other cases. When hold is set to ON, the contacts are closed during holding.
Remote start
Whether the auto calibration is set at ON or OFF, an auto calibration is available by keeping the
remote start input closed for at least 1.5 seconds.
When an auto calibration starts, the measurement screen automatically appears.
Any operation other than “Stop Auto Calibration” (see Item 6.4.2) is
not permitted during auto calibration. “Stop Auto Calibration” cannot be
performed with the key lock to ON. To cancel auto calibration forcedly, set
the key lock to OFF and then execute “Stop Auto Calibration”.
Caution
•
•
Remote start input
Open
Closed (keep at least 1.5 sec.)
Start Time
Cycle
Flow Time
ON/OFF
SUN
1
350
ON
12:00
day
sec
(An example of “CH1: through CH5: enable”, as given in Item 6.2.4 “Auto Calibration
Components”)
Flow time
: Auto calibration
Sunday
12:00
Cycle
Zero
calibration
350sec
350sec
CHI
Span
calibration
350sec
CH2
Span
calibration
CH3
Span
calibration
CH4
Span
calibration
Monday
12:00
Tuesday
12:00
Example
In case where auto calibration is carried out at the above setting.
CH5
Span
calibration
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7days)
Flow time
: 60 to 599 sec
(initial value 300sec)
Setting range