6 - 5
INZ-TN2ZKJ-E
t
In the “Manual Calibration Selection”
screen that appears, select “at once” or
“each” by pressing the
or
key.
When selecting “at once”, the CH (com-
ponents) to be set can be zero-calibrated
at the same time. When selecting “each”,
either of the CH (components) to be
selected is zero-calibrated. After setting,
press the
ENT
key.
Whether “each” or “at once” can be determined for each CH (component).
•Setting “each”
Select the CH (component) on the manual zero calibration screen and then perform
zero calibration.
•Setting “at once”
At a manual zero calibration, zero of CH (components) for which “at once” was selected can
simultaneously be calibrated.
* When the cylinder air or atmospheric air is used for the zero gas, select “At once”.
Example
(
)
ENT
End of Manual Calibration Setting
To close “About ZERO Calibration” setting or to
cancel this mode midway, press the ESC key.
A previous screen will return.
To close “About ZERO Calibration”
ESC
•
When setting all components to “each”:
A single cursor will appear.
Manual Calibration screen
•
When setting all components to “at once”:
Cursors will appear at all components
where “at once” is set.