
6 - 6
INZ-TN2ZKJ-E
6.2.3 Setting of calibration range
This mode is used to set if the range of each CH (component) at the calibration (manual calibration
or auto calibration) should be calibrated with a single range or 2 ranges.
q
During measurement, press the
MODE
key
to display the User mode.
w
Point the cursor to “Setting about Cali-
bration” by pressing the
or
key.
Press the
ENT
key.
e
In the “Setting about Calibration” screen
that appears, point the cursor to “About
Calibration Range” by pressing the
or
key. Press the
ENT
key.
r
In the “Calibration Range CH Selection”
screen that appears, point the cursor to
the CH you want to set by pressing the
or
key. Press the
ENT
key.
(
)
ENT
(
)
ENT
MODE
(
)
ENT