6 - 19
INZ-TN4ZAFE-E
6.6.5 Zero offset setting
Zero offset function is used to eliminate the difference of output due to dirt or deterioration of the
sensor.
Perform zero offset when the off-set value increases up to around 3000 counts.
The zero gas should be delivered when setting offset.
(1) Select “zero off-set” from the menus in the
maintenance mode.
Maintenance
Mode
1. Sensor Input Value
2. Error Log
3. Password Setting
4. Zero off-set
5. Station No. 01
6. Setting of Digital Out
7. Cal. history
8. To Factory Mode
Select operating item
Factory
Mode
H
2
0
(2) Press the ENT key to execute zero off-set.
The displayed value will change to "0".