6 - 13
INZ-TN4ZAFE-E
Calibration
Output hold
Hold extending time
Auto calibration start
Auto calibration end
Time hold set to the gas flow time
Backlight
According to the set time, the backlight in the display unit is turned off. Unless key operation is
performed during the set time period, the backlight is automatically turned off. By pressing an
arbitrary key, the backlight is turned on.
Remote Range
A range can be switched via an external input by setting the Remote Range to ON. If the Remote
Range is set to OFF, the external input becomes invalid.
Opening the input gives the High range, or short-circuiting the input gives the Low range.
Note) In case of 1 range system, this function is overridden.
output Hold
By setting an output hold to ON, an output signal of each channel are held during the calibration
(manual calibration and auto calibration) and for the gas flow time. Regardless of Hold ON/OFF
setting, an output signal can be held via an external input.
(1) Manual calibration
Calibration
Output hold
Press the key to
execute calibration.
Hold extending
time
Time set to gas flow time
(See 6.3 Setting of Auto Calibration)
ENT
Press the key to
flow gas.
ENT
Manual calibration screen
(2) Auto calibration