3.5 Wiring method
Note) After the wiring work, be sure to place the protective cover for the terminal blocks to
assure safety.
The power terminal block and the external I/O terminal blocks are provided on the rear face of the
analyzer. See the following figure.
Wire each terminal by referring to “3.5.1” to “3.5.7”.
Optional terminal block
Standard terminal block
3.5.1 Power supply (standard terminals (1) - (2))
Connect the given power supply to the power terminal, and connect the ground wire to the ground-
ing terminal (standard terminal (3)). Perform class D grounding. Use solderless terminals (for
M3.5) to connect the cables to the terminals.
• Wiring work must be performed with the main power set to OFF to prevent electric shocks.
• Enforce construction of class-D grounding wire by all means.
If the specified grounding construction is neglected, a shock hazard or fault may be caused.
• Wires should be the proper one meeting the ratings of this instrument. If using a wire which can-
not endure the ratings, a fire may occur.
• Be sure to use a power supply of correct rating. Connection of power supply of incorrect rating
may cause fire.
CAUTION