3 - 4
INZ-TN4ZAFE-E
Fig. 3-1 Piping
Sampling gas inlet : Connect the pipe so that zero/span calibration standard gas or measured gas pre-
treated with dehumidification is supplied properly. The gas flow rate should be
kept constant within the range of 0.4L/min ±0.05L/min. (High speed responce:
1L/min±0.05L/min)
Sampling gas outlet : Measured gas is exhausted after measurement.
Connect the pipe so that the gas may escape through the gas outlet into the atmo-
sphere or equivalent.
INLET
OUTLET
Gas inlet
Gas outlet