6 - 12
INZ-TN4ZAFE-E
6.5 Parameter setting
It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be
set are as follows:
* For the maintenace mode, see Item 6.6 Maintenance mode.
• Date
: Sets the date.
• Current Time
: Sets the current day of the week, hour and time.
• Key Lock
: Sets with ON/OFF so that any key operation except the key lock OFF cannot be
performed.
• Backlight
: Automatic OFF time of the backlight in the display unit
• Remote Range
: Sets with ON/OFF whether the Range Selection is made valid or invalid by
external input.
• Output Hold
: Sets whether Calibration Output is held or not.
• Average Value Reset : Resets the average value.
• Response Speed
: Sets the response time of electrical system.
• Average Time
: Sets the moving average time.
• Maintenance mode
: Enters passwords to switch to the Maintenance mode.
Description of setting items
(1) To display the User mode, press the
MODE
key in the measurement mode.
Point the cursor to “Parameter Setting”
by pressing the key. Press the
ENT
key.
To close the "Parameter Setting" screen
or cancel this mode midway, press the
key.
A previous screen will return.
To close Parameter Setting screen
ESC
• Backlight
: 5 to 99 min (Initial value: 10 min)
• Response time
: 1 to 60sec. (Initial value: 3 sec)
• Maintenance mode
: 0000 to 9999 (Initial value: 0000)
Setting range
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
Select an item
with UP/DOWN and ENT
Parameter
Setting
Current Date
2013-07-08
Current Time
MON 16 : 19
Kye Loc
OFF
Back Light
ON 99min
Output Hold
OFF
Response Time
01 sec.
To Maintenance Mode 0000
Select setting item
(2) In the “Parameter Setting” screen that
appears, point the cursor to any item you
want by pressing the key. Press the
ENT
key.