6 - 4
INZ-TN4ZAFE-E
6.2 Alarm setting
(When concentration alarm contact output has been selected)
6.2.1
setting of alarm values
This mode is used to set the upper and lower limit value to provide an alarm output during mea-
surement. Before changing the alarm setting, set the ON/OFF to OFF.
Then set the ON/OFF to On again.
(1) During measurement, press the
MODE
key to
display the User mode.
Point the cursor to “Alarm Setting” by
pressing the key. Press the
ENT
key.
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
Select an item
with UP/DOWN and ENT
Back with ESC
(2) After the Alarm Setting screen has ap-
peared, operate the key until the
cursor is aligned with a desired item and
press the
ENT
key.
“DO3” is an output terminal number.
(3) Enter the setting value pressing the ,
or key. And then press the
ENT
key.
Alarm Setting
Upper
DO3
Lower
DO3
Renge 1 0.900 vol%
Renge 2 03.00 vol%
Renge 1 0.200 vol%
Renge 2 01.00 vol%
Kind of Alarm
Upper+Lower
Hysteresis
00%FS
ON/OFF
OFF
Select an item
with UP/DOWN and ENT
Back with ESC
Alarm Setting
Upper
DO3
Lower
DO3
Kind of Alarm
Upper+Lower
Hysteresis
00%FS
ON/OFF
OFF
Set value
Renge 1 0.900 vol%
Renge 2
0
3.00 vol%
Renge 1 0.200 vol%
Renge 2 01.00 vol%
To close “Alarm Setting” or to cancel this mode
midway, press the key.
A previous screen will return.
To close “Alarm Setting”
ESC