6 - 3
INZ-TN4ZAFE-E
current
Cal. Settings
Cal. Range
Set calibration range
current or both range
Range1
H
2
Range2
0–3 vol%
0–10 vol%
1.28
ZERO Cal.
ENT : Gas floes
ESC : Back
Range1
H
2
Range2
0–3 vol%
0–10 vol%
(4) In the “Calibration Range” screen that
appears, select “Both” or “Current” (for 2
ranges) and press the
ENT
key.
• When selecting “both”, Range 1 and
Range 2 are calibrated together.
• When selecting “Current”, the range
alone displayed is calibrated.
Note) When calibration is performed by the “Both” setting under the normal operating condition,
prepare a span gas cylinder on the normal operating range side. It is recommend to
perform span gas calibration in the normal operating range.
In case of “both” setting
Two cursors will ap-
pear in both ranges
Manual Calibration screen
Select “both” or “current” pressing
by the key.
To close “Setting of Calibration Range” or
to cancel this mode midway, press the key.
A previous screen will return.
To close “Setting of Calibration Range”
ESC