6 - 1
INZ-TN4ZAFE-E
6. sEttInG AnD CALIBRAtIon
6.1 Calibration setting
Calibration setting is made to select the concentration at calibration and calibration operation for the
range.
6.1.1 setting of calibration concentration
Select the concentration of standard gases (zero gas, span gas) used at the time of calibration
following the procedures shown below.
User Mode
Setting about Calibration
Alarm Setting
Setting of Auto Calibration
Changeover of Range
Parameter Setting
Select an item
with UP/DOWN and ENT
Back with ESC
(1) Press the MODE key in measurement
state to display the User Mode screen.
Move the cursor to Calibration Setting by
pressing the key and then press the
ENT
key.
To close “Calibration setting” or terminate the
operation, press the
ESC
key, and you will return to
the previous screen.
To close “Calibration setting”
Cal. Settings
Calibration Value
About Calibration Range
Select an item
with UP/DOWN and ENT
Back with ESC
Cal. Settings
Cal. Value
0 0 . 0 0
RANGE
ZERO
SPAN
H
2
Select setting value
0 0 . 0 0
0–3vol%
0–10vol%
2 . 9 1 0
0 9 . 8 1
(2) In the calibration setting item screen that
appears, move the cursor to “Calibration
Value” by pressing the key, and then
press the
ENT
key.
(3) Select the concentration item you want to
make the setting by pressing the ,
or the
ENT
key.
(4) Enter zero and span calibration gas
concentration by using the key, and
then press the
ENT
key.