![Fuji Electric FCX-AII-VG Series Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/fuji-electric/fcx-aii-vg-series/fcx-aii-vg-series_instruction-manual_2341971048.webp)
45 TN5FCXA2VG-E
M
①
②
③
④
M
M
(Cancel the setting)
(Save the setting)
Menu 2 : Model code (TYPE)
Model code of field device is displayed and changed
(example of differential pressure transmitter).
• Press the key on the screen
j
to display the
model code setting screen (
k
).
• Input alphanumeric characters as required with the
and keys on the screen
k
.
Functions of the keys:
key: To input characters at the cursor position.
(0 to 9, space, A to Z, –)
key: To move the cursor position to the next.
(1
2
3 ...
16
1)
Note)
Characters can be input up 24.
Characters other than numerical characters, capi-
tal letters of the alphabet, space, and “–” are dis-
played as “*.”
Initial six characters are displayed. (The cursor
position is displayed by a vertical bar.)
To display the seventh and following characters,
scroll the characters to the left. (The cursor posi-
tion (far right) is displayed as a number.)
The cursor position is 2 in the example
k
. (“K” is
input as the second character.)
The cursor position is 8 in the example
l
. (“5” is
input as the eighth character.)
• Select whether the type setting is saved on the
screen
m
.
Press the key to save the type setting.
Press the or key to cancel the setting.
* Description of the displays on the first line on the
item name selection screen (
j
)
: Differential pressure transmitter
: Pressure (gauge pressure) transmitter
: Absolute pressure transmitter