3 TN5FCXA2VG-E
ELECTROMAGNETIC COMPATIBILITY
All models FCX-AII series transmitters type FCX-AII VG are in accordance with :.
The harmonized standards :
•
EN 61326-1
(Electrical equipment for measurement, control and laboratory use - EMC requirements)
•
EN 61326-2-3
(Part 2-3 : Particular requirements - Test configuration, operational conditions and performance criteria
for transducers with integrated or remote signal conditioning).
Emission limits : EN 61326-1
Frequency range (MHz)
Limits
Basic standard
30 to 230
40 dB (µV/m) quasi peak measured at 3m distance
Passed
230 to 1000
47 dB (µV/m) quasi peak measured at 3 m distance
Immunity
Phenomenon
Test value
Basic standard
Performance criteria
Electrostatic Discharge (ESD)
4 kV (Contact)
EN 61000-4-2
B
8 kV (Air)
IEC 61000-4-2
Electromagnetic Field
10 V/m (80 to 10 GHz)
EN 61000-4-3
3 V/m (1.4 to 2.0 GHz)
IEC 61000-4-3
A
1 V/m (2.0 GHz to 2.7 GHz)
Rated power frequency
30 A/m (50 Hz, 60 Hz)
EN 61000-4-8
A
Magnetic field
IEC 61000-4-8
Burst
2 kV (5/50 ns, 5 kHz)
EN 61000-4-4
B
IEC 61000-4-4
Surge
1 kV line to line
EN 61000-4-5
B
2 kV line to ground
IEC 61000-4-5
Conducted RF Disturbances
3 V (150 kHz to 80 MHZ)
EN 61000-4-6
A
IEC 61000-4-6
Performance criteria (A & B): according to IEC 61326-1:2013, § 6.4.