Fault description
Cause
Remedy
While setting the camera in the
"Measurement settings" configuration
item, a black image is shown.
Camera not connected
correctly
Check the camera connections! Start the soft-
ware again! (see
No commands executed by the A28!
USB cable not connected
correctly
Check USB cable and connect it properly to the
device and PC! Start the software again!
No settings can be made in the feeder
'setup'
USB cable not connected
correctly
Check USB cable and connect it properly to the
device and PC! Start the software again!
Dots are visible in the same places in
all images
Soiled windows, lens or
camera lens
Clean these parts as described in
Dark corners are visible in the preview
image while setting up the camera
Camera settings not cor-
rectly adjusted
Optimise camera settings (gain, flash)