7.
If necessary and depending on the setting in the SOP, the intensity
of the feeder can vary. If using the wet dispersion unit, gradually
add sample material to the sample bath. You can start the meas-
urement once the beam absorption is correct.
If using the wet version of the ImageSizer, make
sure to wait a few seconds each time you add
sample material to allow the sample material to
be distributed in the measuring circuit. In this
way, you can add add the optimal amount of
sample material.
The current beam absorption is displayed in the
lower line as ‘good, high and low density’ .
8.
Once the cancellation criteria "Image count" or "Particle count"
are reached, the measurement is finished and you can display the
evaluation by clicking on "Stop" and "Close".
9.
When the measurement is finished, if using the dry version,
remove the sample from the collecting vessel and clean the parts
that come into contact with the samples before you begin to
measure other materials. The wet dispersion unit must be thor-
oughly rinsed after a measuring process or a series of measure-
ments.