7.6.3 Emptying the system
1.
Check whether the hose at the "Liquid in" connection is connected
via the pressure controller to your water connection or a liquid
storage container.
2.
Open the tap of the water connection or the storage container.
3.
Start ISS.
4.
Open SOP management and insert a new SOP. Insert the ‘Empty’
command.
5.
Then press the button with the triangle on the right side of the
command to execute the command.
7.6.4 Removing the wet measuring cell
Although the measuring circuit should be completely without liquid, it is
possible that liquid residues remain in the measuring cell itself. As a pre-
caution use a paper towel or similar to immediately collect liquid flowing
out.
1.
Lift the measuring cell drawer out of the measuring unit and place
it on a prepared piece of paper towel.
Remove the 4 screws which close the measuring cell.
NOTICE!
Ensure that the rear cover and the measuring cell
glass can not fall out.
2.
The rear flange falls out of the casing when the 4 screws are
released. Take care not to scratch the rear measuring cell glass
during disassembly. The front flange can now be removed just as
carefully.
3.
All parts can now be cleaned. Extreme care must be taken when
cleaning the measuring cell glasses. To do this, proceed as
described in
Chapter 7.6.5 ‘Cleaning the measuring cell glasses’
7.6.5 Cleaning the measuring cell glasses
Great care must be taken when cleaning the measuring
cell glasses. The glasses must only be touched by hand
at the edges.
Cleaning
- 49 -