5.4 Fritsch test powder F16
You can check the precision and accuracy of your device
by measuring the Fritsch test powder at least once a
week, depending on how often you use the device. This
will indicate contamination or malfunctions in your
system.
Measure the supplied Fritsch test powder F16 with the ANALYSETTE 28
ImageSizer as follows:
1.
Set the device to I (On) with the main switch.
2.
Start the Image Sizing software.
3.
Create a new folder in the archive.
4.
A "New Project Case" must be created In the newly created folder.
5.
In this project case you can select the SOP "F16 Standard measure"
by clicking on "New measurement".
SOP refers to a Standard Operating Procedure
used in our software. In an SOP, work steps are
pre-defined and processed with a single key‐
stroke.
Ensure that you select the SOP which is suitable
for your device.
6.
The measuring window with the live image, camera and flash set-
tings opens.
7.
If necessary and depending on the setting in the SOP, the intensity
of the feeder can vary. Once the beam absorption is correct, you
can start the measurement.
The current beam absorption is displayed in the
lower line as ‘good, high and low density’ .
8.
Once the cancellation criteria "Image count" or "Particle count"
are reached, the measurement is finished and you can display the
evaluation by clicking on "Stop" and "Close.
9.
When the measurement is finished, remove the sample from the
collecting vessel and clean the parts that come into contact with
the samples before you begin to measure other materials.