13
2.3.4.
Stopping Data Recording
Click
Stop
to stop recording. In Stop mode, MicroLite keeps all recorded data but
does not record new data. Use this mode to save battery power.
2.3.5.
Temperature Sensor Calibration
A new MicroLite comes fully calibrated. After a long period of use however, you may
want to recalibrate the Temperature sensor.
The calibration is made directly to the MicroLite firmware i.e. the calibrated data is
displayed on the logger LCD, not just in the software following data download.
Calibration must be carried out while MicroLite is connected to the PC.
Calibration Password
To prevent accidental change of the calibration, the calibration procedure is protected by
a password.
The default password is: 1234.
To change the password:
1. Click
Logger
on the main menu, and then click
Calibration
.
2. Click
Change Password
to open the
Change Password
dialog box.
3. Enter the current password in the
Current Password
field.
4. Enter the new password in the
New Password
field.
5. Enter the new password a second time in the
Confirm New Password
field to
confirm your new password.
6. Click
OK
.
Note:
The password must include at least 4 characters and is
case sensitive.
Temperature Calibration
1. Connect the MicroLite to the PC.
2. Click
Logger
on the main menu, and then click
Calibration
.
3. Enter the calibration password, and then click
OK
.
4. The Calibration dialog will open.
Figure 5: Calibration dialog box