12
Click the minutes (mm) and type 5 or use the arrows, and finally,
click the seconds (ss) and type 30 or select 30 using the arrows.
Recording Time
Displays the total recording time according to the logging interval
selected.
Timer run
Select the
Timer run
check box if you want MicroLite to start
recording at a predetermined time. This option is convenient if
you are using several MicroLite at once and want them to all
begin logging at the same time.
Use the time and date selectors to set the start time.
Cyclic run
In
Cyclic run
mode, the MicroLite overwrites the old
measurements (starting with the oldest recorded data) once the
MicroLite’s memory is full. Click the
Cyclic run
check box to
operate in this mode.
If the check box is unselected, MicroLite will operate in
Normal
run
mode and will stop recording when the memory is full.
Push to run
In
Push to run
mode, the MicroLite will only start recording data
when you use the MicroLite magnet to close the reed relay on the
MicroLite’s circuit board, thereby causing the logger to start
recording data. This is convenient for when you wish to start
recording data soon (but not immediately) following setup, or at
an unspecified time.
2.3.2.
Completing the Setup
Click
Send Setup
to send the new settings to the MicroLite and begin recording. This
will complete the setup. Click
Cancel
if you do not wish to change the setup at this
stage.
Note
: The
Send Setup
command erases all existing data in the MicroLite.
If you select
Timer run
mode, MicroLite will wait in standby mode, displaying
tRUN
. It
will begin recording at the specified time.
If you select
Push to Run
mode, MicroLite will also wait in standby mode, displaying
PUSH
until the magnet is used to start the logger.
2.3.3.
Starting Data Recording
Click
Run
every time you want to start a new recording. The
Run
command erases
all previous data in the MicroLite memory and begins recording.
Note:
If you setup the MicroLite using the
Setup
command, it will automatically begin
recording, and you don’t have to click
Run
.
www.esis.com.au