![Formulatrix NT8 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/formulatrix/nt8/nt8_user-manual_2320287069.webp)
65
NT8 User's Guide | Chapter 15: System Maintenance
Removing the NT8’s Covering
1.
Make sure the NT8 is powered off and unplugged.
2.
Open the hood to expose the NT8’s head and deck.
3.
Using your hands, gently move the head to the center of the NT8.
4.
Select an appropriately sized hex key from the hex key set, and remove the four
screws that attach the dark gray cover to the NT8’s head. There are two screws on the
right and two on the left.
5.
Gently remove the cover to expose the inner workings of the NT8’s head.
Lubricating the Plate Copy Plungers
1.
Plug the NT8 into a power supply and power it on.
2.
Start the NT8 software and wait for it to connect to the NT8 hardware.
3.
Click the
Programmatic Mode
Tab .
4.
Go to the Control Manual Motion and click the UP arrow below to C by using a
medium speed. The plate copy plungers will move up, exposing eight metal cylinders.
5.
Using the paintbrush, lightly but evenly apply grease to the surface of each of the 8
metal cylinders. Try to reach the back side of the cylinders as well.
6.
Back in the software, click the
Load Script
button .
7.
Locate and open the
PCTesting.msq
script.