NT8 User's Guide | Chapter 6: Initial Setup
Attaching LCP Attachments
An LCP dispensing attachment is available for the NT8 that can dispense LCP drops of 50 -
100 nL. The dispenser uses an 8 mm barrel syringe. A sensor ensures the syringe is properly
mounted before the NT8 begins dispensing drops. To install the LCP attachments, you'll need
to locate the LCP syringe, which are included in the NT8 LCP accessory kit.
Prerequisites
●
You want to perform the LCP drop dispense.
●
An LCP syringe.
Installing the LCP Syringe
Before you perform the LCP drop dispense, you'll need to install the parts. To install the LCP
syringe, take the following steps:
1.
Install the LCP syringe by inserting it into the LCP holder slot.
2.
Push the LCP syringe to make sure it is firmly attached to the syringe holder.
3.
After the plunger has been inserted, tighten the thumb screw by turning it clockwise.
Note:
You can adjust the lever speed and height to fit the LCP syringe by clicking on the
toolbar and moving the lever button on the NT8's head up and down.