viii TIMS MVP
™
Setup and Configuration Guide
Table 2
–
Guidance and MANUFACTURER’S declaration –
electromagnetic IMMUNITY
–
for all ME EQUIPMENT and ME
SYSTEMS
(see 5.2.2.1 f)
Guidance and manufacturer’s declaration
—
electromagnetic immunity
The TIMS MVP
™
System is intended for use in the electromagnetic environment
specified below. The customer or the user of the TIMS MVP
™
System should assure that
it is used in such an environment.
When the TIMS MVP
™
System is exposed to electromagnetic (EM) disturbances, the
system may show abnormal behavior. For instance, the measured traces may be
become obscured, the software may crash or, in case of very high-level transient
voltage events like for instance ESD, parts of the system may even become defective.
Immunity test
IEC 60601 test
level
Compliance level Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for
input/output lines
±2 kV for AC and
DC powerlines
±1 kV for
input/output
lines
±2 kV for AC and
DC powerlines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 differential
mode
±2 kV common
mode
±1 differential
mode
±2 kV common
mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
0 % UT
(100 % dip in UT)
for 0,5 cycle at
0°,45°,90
°,135°1
80°,225°,270°
and 315°
70% UT
(30% dip in UT)
for 25 cycles
70% UT
(30% dip in UT for
25 cycles)
0 % UT
(100 % dip in UT)
for 0,5 cycle at
0°,45°,90
°,135°1
80°,225°,270°
and 315°
70% UT
(30% dip in UT)
for 25 cycles
0% UT
(100% dip in UT
for 5 sec)
Mains power quality should
be that of a typical
commercial or hospital
environment, If the user of
the TIMS MVP
™
System
requires continued
operation during power
mains interruptions, it is
recommended that the
TIMS MVP
™
System be
powered from an
uninterruptible power
supply or a battery.