x TIMS MVP
™
Setup and Configuration Guide
Table 4
–
Guidance and MANUFACTURER’S declaration –
electromagnetic IMMUNITY
–
for ME EQUIPMENT and ME SYSTEMS
that are not LIFE-SUPPORTING
(see 5.2.2.2)
Guidance and manufacturer’s declaration
—
electromagnetic immunity
The TIMS MVP
™
System is intended for use in the electromagnetic environment specified below. The
customer or the user of the TIMS MVP
™
System should assure that it is used in such an environment.
When the TIMS MVP
™
System is exposed to electromagnetic (EM) disturbances, the system may show
abnormal behavior. For instance, the measured traces may become obscured, the software may crash
or, in case of very high-level transient voltage events like for instance ESD, parts of the system may
even become defective.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment
—
guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the TIMS MVP
™
System, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Conducted RF
IEC61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
6 Vrms for ISM
bands
3 V/m
80 MHz to 2.7 GHz
27 V/m
385 MHz
28 V/m
450 MHz
9 V/m
710/745/780 MHz
28 V/m
810/870/930 MHz
28 V/m
1720/1845/1970
MHz
28 V/m
2450 MHz
9 V/m
5240/5500/5785
MHz
3 V/m
80 MHz to 800 MHz
800 MHz to 2,7 GHz
where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
The TIMS MVP
™
System is fairly sensitive to
conducted and radiated RF. Disturbance of the
TIMS MVP
™
System trace is possible at and
below the specified test level. It may be
necessary to relocate the TIMS MVP
™
System
or to apply shielding.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.