![Fona Aquamarine Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/fona/aquamarine/aquamarine_operating-instructions-manual_2315301064.webp)
64
Aquamarine
Accustic signal
The laser diode has reached the limit temperature, max. 4 minutes of
continous work at maximum power.
● Wait until the laser diode has cooled down.
If you cannot solve the malfunctioning, switch off the laser and contact the
manufacturer, your local dental dealer or your authorized service center.
8.6
Technical support, repair and testing
The manufacturer provides technical information on the repair of
individual components only to authorized dealers and only after
conducting an advanced training course for their technical personnel.
Please contact your local dental depot or an authorized Customer Service
Department for technical support.
The Aquamarine may be sent in for repair or for safety inspection only in
its original packaging, including all accessories. Disinfect the Aquamarine
and sterilize the accessories according to the relevant instructions for use
before shipping them.
8.7
Replacing the rechargeable battery of the
control unit
If the rechargeable battery does not load more than 30% even by
charging it overnight, the battery pack should be replaced.
1. Disconnect the power supply.
2. Unlock the quick release fastener (A).
3. Remove the battery pack (B).
WARNING
Do not touch the battery pins and the patient simultaneously.
Содержание Aquamarine