TEMPERATURE INVALID
Invalid temperature (< -10°C or > 150°C) was detacted.
1. step: Create valid environmental conditions.
If error message remains: Software update
If error message still remains: Exchange the handpiece
480
CURRENT TO HIGH
Current value too high. The current value measured by the
current sensor is 5% higher as the expected value according
to the calibration / characteristic curve.
1. step: Software update
If error message remains: Exchange the handpiece.
481
CURRENT TO LOW
Current value to low. The current value measured by the
current sensor is 5% lower as the expected value according
to the calibration / characteristic curve.
1. step: Software update
If error message remains: Exchange the handpiece.
490
BASE DEVICE TIMEOUT
The communication monitoring on the side of the handpiece
has not received a message from the communication
monitoring of the control unit within the requested timeframe.
Short-term solution: Restart of control unit and handpiece
Long-term solution: Software update
500
DOOR SWITCH OPEN
Open door switch. The door switch contact of the control unit
is open.
Close the contact.
501
EMERGENCY STOP PRESSED
The emergency stop button was pressed.
Reset the emergency button:
600
INTERNAL SOFTWARE ERROR
Internal software error
Short-term solution: Restart of control unit and handpiece
Long-term solution: Software update
Error code
Error message
Cause / Solution